Bair Hugger Model 505 Service Manual
14
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Over Heat Alarm Testing
5. Allow the warming unit to cycle 2 times more after the first
satisfactory reading. When the alarm activates, note the
alarm temperature read-out from the test kit. See Table 1
for specified limits.
NOTE: The first over heat cycle temperature may be slightly
higher when started from a cold start.
6. If the results of the test did not meet specified limits,
adjust the over heat thermostat. See the Over Heat Alarm
Thermostat Adjustment section.
7. If the results of the test did meet specified limits, turn the
warming unit to STANDBY, and disconnect the warming
unit from the power source.
8. Remove the hose end from the port on the Temperature
Test Unit.
9. Retighten the screw on the over heat alarm test panel.
10. Record the maintenance action taken and hours of
operation (read from the hour meter) in the Maintenance
Log section.
Table 1. Over Heat Specification
High Temperature Thermostat Model 505
The thermostat interrupts power to the heater and activates
the alarm at a preset high temperature of 53° ± 3°C
(127.4° ± 5.4°F) or less at the end of the hose, as determined
by the Temperature Test Unit Model 22110.