Bair Hugger Model 505 Service Manual
20
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Normal Operating Temperature Adjustment
3. Insert the slotted screwdriver in the potentiometer access
hole. You must access the potentiometer blindly, so carefully
move the screwdriver until it fits into the slot of the
potentiometer.
• Ifthetestedoperatingtemperaturewasbelow specified
limits, turn the screwdriver CLOCKWISE to increase
the operating temperature.
• Ifthetestedoperatingtemperaturewasabove specified
limits, turn the screwdriver COUNTERCLOCKWISE
to lower the operating temperature.
4. Adjust the potentiometer until the temperature read-out
meets the limits specified in Table 2 for the HIGH
temperature setting (correlated to ambient temperature as in
Figure 8).
5. Verify that the MED and then LOW temperature settings
are also within specified limits. If the HIGH range is
properly set, and MED and/or LOW do not meet specs, call
Technical Service.
6. Place the warming unit in Standby and remove the hose
from the port on the Temperature Test Unit.
7. Record the maintenance action taken and hours of operation
(read from the hour meter) in the Maintenance Log section.
Figure 9. Location of
Operating Temperature
Adjustment Potentiometer
Operating temperature
potentiometer access