Bair Hugger Security Systems Service Manual Page 15

  • Download
  • Add to my manuals
  • Print
  • Page
    / 37
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 14
11
Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Accessing the Alternative Modes
To access the Alternative modes, you must press the Alternative Mode button that is located on the front
panel of the temperature management unit but is unlabeled. Press the text on the Bair Hugger logo when
the instructions direct you to press the Alternative Mode button.
Figure 2. Location of Alternative Mode button
Alternative
Mode button
Over-temperature
Detection Test
AltMode4
System Information
AltMode3
Fault Code Log
AltMode2
Calibration
AltMode1
To enter one of the Alternative modes, press the Alternative Mode button in combination with one of the
operating mode buttons, also located on the front panel. Figure 2 indicates which keys must be pressed to
enter each of the 5 Alternative modes. When the correct combination of keys is pressed, the display reads
“Hold for AltModeX,” where “X” represents the selected Alternative mode from the list above.
Exiting an Alternative Mode
Exit the Alternative modes by pressing the Alternative Mode button or by waiting for 60 seconds. You will
need to access the Alternative modes for several of the service procedures described in this manual.
Page view 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 36 37

Comments to this Manuals

No comments