Bair Hugger Model 750 Temperature Management Unit Service Manual
14
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Testing the Over-Temperature Detection System
warning
Do not perform the over-temperature detection system test while the temperature management unit is being
used for temperature management therapy. ermal injury may result.
caution
Connect each warming unit being tested to a separate power source.
Service frequency
Every 6 months or 500 hours of use, or whenever service is performed.
Tools/equipment
Model 22110 Temperature Test Unit
Method
: If continuing from previous section (Calibrating the Operating Temperature) skip to step 4.
: If the temperature management unit has been stored at a temperature lower than 20°C, allow it to
reach ambient temperature before beginning the test.
1. Connect the temperature management unit to a grounded power source. e unit performs a self-test
and automatically proceeds to Standby mode
2. Connect the temperature test kit to the hose of the temperature management unit.
3. Press the Ambient button and allow the temperature management unit to run for ve minutes.
4. Press the Alternative Mode button (See Figure 2) and the 43 °C button simultaneously for three
seconds. e temperature management unit is now in a Forced Over-temperature mode. In this mode,
the temperature management unit performs the following tasks:
• Blower activates.
• Heater elements continuously operate at full power.
• 43°C indicator light ashes.
• Over-temperature timer activates.
• “OT Test” ashes in the alphanumeric display.