Bair Hugger Model 750 Temperature Management Unit Service Manual
26
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Fault Code Possible Cause Solution
Cautionary Errors
Fault condition while unit is in use
FC-001 Hose end at blanket end
occluded.
Disconnect from power source.
Remove hose occlusion. Position
hose end correctly into hose card
(refer to the operator’s manual).
Wait 30-60 seconds before
reconnecting to the main power
source. Reconnect to power
source. If no fault occurs, unit is
ready for use.
FC-002
FC-003
FC-004
FC-005
FC-006
FC-007
Hose end sensor failure at
blanket end. Faulty elbow
sensor.
Moisture.
Intermittent sensor failure.
High heat.
Disconnect from power source.
Allow unit to cool. Wait 30-
60seconds before reconnecting
to the main power source.
Reconnect to power source. If no
fault occurs, unit is ready to use.