Bair Hugger Model 775 Temperature Management Unit Service Manual
8
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Operating Modes
e Model 775 temperature management unit has 4 operating modes: Ambient, 32°C, 38°C, 43°C
To select the 32°C, 38°C, 43°C, or Ambient mode, press the corresponding button� e temperature
management unit will operate within the specied temperature range, shown in the table below
Temperature Mode Temperature of Delivered Air
32°C 32°C ± 1�5°C
38°C 38°C ± 1�5°C
43°C 43°C ± 15°C
Ambient Unit will supply air that is slightly warmer than room-temperature
Aer you select a temperature mode, the following events occur:
• Corresponding indicator light illuminates�
• Blower operates�
• Heater activates except in Ambient mode
• Temperature mode timer and hour meter activate
• Temperature at the blanket-end of the hose appears in the alphanumeric display
• Temp in Range indicator light illuminates when the temperature at the blanket end of the hose is
within ±1�5°C of the selected setting; this indicator light does not illuminate in the Ambient mode�
Airflow Modes
e Model 775 has two preset fan speed settings: a default or high fan setting ( )and a reduced or low
fan setting ( )� e Model 775 will reset/default to the high fan setting each time the unit is powered up
or Standby is selected� e low fan setting ( ) may be pre-selected while in Standby prior to selecting the
desired temperature�
Standby Mode
To place the unit in Standby mode, press the Standby button� When in Standby mode, the following events
occur:
• Standby indicator light illuminates�
• Blower and heater turn o�
• Alphanumeric display deactivates�
• Temperature mode timer pauses
• Alarm and fault detection functions remain active�