Bair Hugger Security Systems Service Manual Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 35
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
Bair Hugger Model 775 Temperature Management Unit Service Manual
10
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Alternative Modes
e Bair Hugger Model 775 temperature management unit has 5 alternative modes that allow access to the
service and calibration utilities of the unit (see Figure 2 on page 11)�
Alternative Modes Description
AltMode0 Hour Meter
AltMode1 Calibration
AltMode2 Fault Code Log
AltMode3 System Information
AltMode4 Over-temperature Detection Test
e Hour Meter mode displays the cumulative amount of operating time (in Ambient, 32°C, 38°C, and 43°C
modes) the temperature management unit has been in use� e time is displayed in whole hours only
e Calibration mode allows a technician to check and adjust the hose end sensor calibration� is should
be done as part of routine maintenance and aer servicing the temperature management unit or replacing
the hose� e hose end sensor temperature can be altered a maximum of ±3�0°C by adjusting the calibration
oset value� In the Calibration mode, the temperature management unit operates in the Ambient temperature
mode, and the alphanumeric display shows the current hose end sensor temperature and calibration oset
with 0�1°C precision�
e Fault Code Log mode sequentially displays the last 5 fault codes stored in system memory� Please refer to
the "Fault Code Table" on page17 for more information about a particular fault code
e System Information mode allows the technician to view important system information including the
rmware, system identication, serial port baud rate, 5V power supply status, and the control board serial
number�
e Over-temperature Detection Test mode allows the user to test the operation of the independent over-
temperature shutdown system
Page view 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 34 35

Comments to this Manuals

No comments