Bair Hugger Security Systems Service Manual Page 23

  • Download
  • Add to my manuals
  • Print
  • Page
    / 35
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 22
19
Bair Hugger Model 775 Temperature Management Unit Service Manual
English 202526A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Replacing the Air Filter
Service frequency
Every 12 months or 500 hours of use�
Tools/equipment
• Replacement air lter
• #2 Phillips screwdriver
WARNING
Do not attempt to clean the air lter as it may contain hazardous microorganisms� Discard the lter in a
manner consistent with institutional protocol for biohazardous material
Method
1� Disconnect the temperature management unit from the grounded power source�
2� Turn the temperature management unit upside down�
3� Remove the 2 screws located in the louvered lter cover
4� Li the lter cover o the temperature management unit
5� Remove the air lter and discard it�
6� Place the new lter in the lter compartment with the gasket toward the lter ledge�
7� Replace the lter cover with the louvers facing the back of the temperature management unit�
8� Replace the 2 screws in the lter cover
Page view 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 34 35

Comments to this Manuals

No comments