Bair Hugger Model 775 Temperature Management Unit Service Manual
24
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Fault Code Possible Cause Solution
Cautionary Errors
Fault condition while unit is in use
FC-001 Hose end at patient end
occluded�
Disconnect from power source�
Remove hose occlusion� Position
hose end correctly into hose card
(refer to the operator’s manual)�
Wait 30-60 seconds before
reconnecting to the main power
source� If no fault occurs, unit is
ready for use�
FC-002
FC-003
FC-004
FC-005
FC-006
FC-007
Hose end sensor failure at
patient end� Faulty elbow
sensor�
Moisture
Intermittent sensor failure�
High heat�
Disconnect from power source�
Allow unit to cool� Wait 30-60
seconds before reconnecting to
the main power source� If no fault
occurs, unit is ready to use