Specifications
Physical Characteristics
 13" high x 13" deep x 14" wide
33 cm high x 33 cm deep x 36 cm wide
 16 lb (7.3kg)
   53 dBA (High fan setting)
48 dBA (Low fan setting)
 Flexible, compatible with the Bair Hugger brand 241
blood/fluid warming system.
  High efficiency 0.2 µm air filter
   Every 12 months or 500 hours of use.
 Can be clamped to an IV pole, placed on a hard
surface, or mounted to the rolling stand accessory.
Temperature Characteristics
   Temperature: 1C-25°C
Humidity: Max relative humidity 80% (up to
31°C) decreasing linearly to 50%
relative humidity at 40°C
Altitude: Max 2,000m
  Electronically controlled.
  High fan setting:
1600 BTU/hr (average), 470 W (average)
Low fan setting:
1330 BTU/hr (average), 390 W (average)
 Average temperatures at the end of the hose:
HIGH: 43° ± 1.5°C 109.4° ± 2.7°F
MED: 38° ± 1.5°C 100.4° ± 2.7°F
LOW: 32° ± 1.5°C 89.6° ± 2.7°F
     2 - 5 minutes (dependent on blanket model)
Time required for the contact surface tempaerature to
heat up from 23 ± 2°C to 37°C (73 ± 2°F to 99°F)
/  -20°C to 45°C (-4°F to 113°F)
Store all components in a cool, dry place when not in
use.
Safety System
 Independent electronic circuit; thermal cutoff shuts
the heater OFF to ensure hose end air remains below
56°C (5C ± 3°C typical); back-up over-temperature
detection at hose inlet.
  Over-temperature (≤56°C, 53°C ± 3°C typical): red
Over-temp indicator light flashes, alarm sounds,
heater and blower shut down, operating indicator
lights turn OFF, control panel becomes unresponsive.
Fault: amber Fault indicator light flashes, alarm
sounds.
  Dual input fused lines.